Medical Device Standards Order (Standards for Medical Devices Required to be Sterile) 2008

Administered by Department of Health, Disability and Ageing

Legislation au F2008L04336 Not in force Legislative Instrument

Legislation content

 

EXPLANATORY STATEMENT

 

Subject:  MEDICAL DEVICE STANDARDS ORDER (STANDARDS FOR MEDICAL DEVICES REQUIRED TO BE STERILE) 2008

 

Section 41CB, Therapeutic Goods Act 1989

 

OUTLINE

 

Medical Device Standards Order (Standards for Medical Devices Required to be Sterile) 2008 (MDSO (Sterile Devices) 2008) is an Order made by the delegate of the Minister for Health and Ageing under section 41CB of the Therapeutic Goods Act 1989 (the Act).

 

MDSO (Sterile Devices) 2008 revokes and replaces Medical Device Standards Order No. 3 – Medical Devices Required to be Sterile (MDSO 3) that was made on 20 February 2003 and commenced upon its gazettal on 5 March 2003.  MDSO 3 of 2003 specified particular standards as medical device standards for medical devices that are required to be sterile, whether the device is to be sterilised by the manufacturer prior to release or to be supplied in a non-sterile state but packaged in such a way that it can be sterilised at a later stage following supply.

 

MDSO (Sterile Devices) 2008 introduces new standards for quality assurance techniques for the manufacture of kinds of medical devices that are intended by the manufacturer to be supplied in a sterile state.  The new standards are based on the following updated standards:

         EN 556-2: 2003 Sterilization of medical devices – Requirements for medical devices to be designated “STERILE” – Part 2: Requirements for aseptically processed medical devices;

         EN ISO 11607-1: 2006 Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems;

         EN ISO 11607-2: 2006 Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming, sealing and assembly processes.

         EN ISO 11135-1: 2007 Sterilization of health care products – Ethylene Oxide – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices;

         AS/NZS ISO 11137-1: 2006 which is identical to ISO 11137-1: 2006 Sterilization of health care products – Radiation – Part 1: Requirements for validation and routine control – Radiation sterilization;

         AS/NZS ISO 11137-2: 2006 Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose;

         AS/NZS ISO 11137-3: 2006 Sterilization of health care products – Radiation – Part 3: Guidance on dosimetric aspects;

         EN ISO 17665-1: 2006 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices;

         EN ISO 11737-1: 2006 Sterilization of medical devices – Microbiological methods – Part 1: Determination of a population of micro-organisms on products

         EN ISO 11737-2: 2000 Sterilization of medical devices – Microbiological methods – Part 2: Tests of sterility performed in the validation of a sterilization process

         ISO 13408-1: 2008 Aseptic processing of health care products – Part 1: General requirements

         ISO 13408-2: 2003 Aseptic processing of health care products – Part 2: Filtration

         ISO 13408-3: 2006 Aseptic processing of health care products – Part 3: Lyophilization

         ISO 13408-4: 2005 Aseptic processing of health care products – Part 4: Clean-in-place technologies

         ISO 13408-5: 2006 Aseptic processing of health care products – Part 5: Sterilization in place

         ISO 13408-6: 2005 Aseptic processing of health care products – Part 6: Isolator systems

         ISO 14937: 2000 Sterilization of health care products – General requirements for characterization of a sterilizing agent and the development of routine control of a sterilization process for medical devices

         EN ISO 17664: 2004  Sterilization of medical devices – Information to be provided by the manufacturer for the processing of resterilizable medical devices

 

Electronic or hard copies of all of the above AS and ISO standards can be purchased on-line from SAI Global Limited, which is accessible at the following website: http://www.saiglobal.com.

 

MDSO (Sterile Devices) 2008 was signed by the delegate of the Minister on 14 November 2008 and commenced on the day after it was registered on the Federal Register of Legislative Instruments.

 

BACKGROUND

 

The Act provides for the establishment and maintenance of a national system of controls relating to the quality, safety and efficacy of therapeutic goods that are used in Australia or exported from Australia.

 

Section 41CB of the Act authorises the Minister, or the Minister’s delegate, by written instrument called an Order, to determine medical device standards for kinds of medical devices set out in the Order and to determine that medical devices of those kinds that comply with the standard are to be treated as complying with those parts of the essential principles specified in the standard. 

 

The essential principles set out the requirements relating to the safety and performance characteristics of medical devices that must be complied with before a device can be imported, supplied or exported.  Compliance with applicable medical device standards is not required, but it is one way to establish compliance with the essential principles. If a manufacturer chooses to apply a medical device standard set out in the Order, and this is applied correctly, the device is presumed to comply with the parts of the essential principles set out in the Order (section 41BH of the Act).

 

MDSO (Sterile Devices) 2008 determines medical device standards by reference to published standards in three areas: qualification of a device as "sterile", packaging, and validation of sterilisation processes. It is essential, as determined by the Order, that all three aspects are applied in combination for the final level of sterility to be achieved.

 

The requirements for devices that are to be supplied "sterile" are set out in Schedule 1 of MDSO (Sterile Devices) 2008 and the requirements for devices which are intended by the manufacturer to be sterilised before they are supplied are set out in Schedule 2 of MDSO (Sterile Devices) 2008.

 

 

CHANGES TO STANDARDS

 

International ‘ISO’ and ‘EN’ device standards are living documents that are developed and constantly being updated by groups of international experts.  Australian representatives are involved in some of these committees.  There is extensive consultation on the ISO and EN standards during their development and subsequent review.  With both industry and the TGA seeking to optimise Australia’s position in the global device market, it is imperative that Australia’s standards do not fall out of step with the international market.

 

Updates to international device standards reflect changes to international best practice as well as the emergence of new technologies and new manufacturing procedures.  Where relevant the latest international standards are adopted by leading regulators including Australia, Europe, the USA and Canada.

 

Australia will very quickly fall behind if it fails to adopt the latest international standards.  In the longer term not keeping up with changes to the relevant international standards will lead to a unique regulatory system in Australia, making regulatory compliance more difficult and costly for importers and/or exporters of medical devices into Australia.  The updated standards referenced in MDSO (Sterile Devices) 2008 reflect the current relevant international standards for sterile devices.

 

CONSULTATION

 

Key industry stakeholders including the Medical Industry Association of Australia (now the Medical Technology Association of Australia), the Australian Dental Industry Association and AusBiotech were consulted on the adoption of an updated version of MDSO 3 (the updated MDSO 3) during January-February 2008.  Industry groups supported the adoption of these standards. 

 

The updated MDSO 3 referenced ISO 13408-1: 1998 which in June 2008 was updated to ISO 13408-1: 2008 Aseptic processing of health care products – Part 1: General requirements.  The TGA therefore further updated this Order to also reference ISO 13408-1:2008, with a proposal to phase out ISO 13408-1: 1998, in September 2009, in line with the European timeframes.  A further round of consultation on this proposal took place in July-August 2008.  Industry groups supported the adoption of the updated standard.

 

REGULATION IMPACT STATEMENT

 

Compliance with the proposed medical device standards is voluntary and members of industry may choose alternative means to demonstrate compliance with the Essential Principles.  All stakeholders, including industry and Standards Australia have been consulted during the development of the proposed new regulatory system for medical devices.  There was overall support for the adoption of international standards.  The Office of Regulation Review assessed the proposal for voluntary standards and, as it is not prohibitive either in terms of costs or time delays, the proposal is considered to be non-regulatory and as such a Regulatory Impact Statement is not required.

 

APPLICATION OF THE LEGISLATIVE INSTRUMENTS ACT 2003 (THE LIA)

 

Under paragraph 6(d)(i) of the LIA, an instrument is a legislative instrument for the purposes of the LIA if it is declared to be a disallowable instrument under legislation in force before the commencement of the LIA.  This determination is a legislative instrument and it is subject to tabling and disallowance in the Parliament under sections 38 and 42 of the LIA, respectively.

 

Overview

The Medical Device Standards Order (Standards for Medical Devices Required to be Sterile) 2008 (MDSO (Sterile Devices) 2008) was enacted to ensure that medical devices in Australia meet internationally recognised standards for sterility. This Order, made under section 41CB of the Therapeutic Goods Act 1989, replaces the previous Medical Device Standards Order No. 3 – Medical Devices Required to be Sterile, which had been in place since 2003. It establishes new standards based on updated international standards to enhance the quality assurance techniques for manufacturing sterile medical devices. The Order was introduced to address the need for Australia to remain aligned with international best practices and to avoid regulatory disparities that could complicate compliance for medical device importers and exporters. The Minister for Health and Ageing, through the delegate, signed the Order on 14 November 2008, and it commenced on the day after its registration on the Federal Register of Legislative Instruments. The objective is to ensure that the standards for sterile medical devices reflect the latest technological advancements and manufacturing procedures, thereby maintaining the safety and efficacy of therapeutic goods in Australia.

Scope and Application

The Medical Device Standards Order (Standards for Medical Devices Required to be Sterile) 2008 applies to medical devices that are required to be sterile, whether the device is to be sterilised by the manufacturer prior to release or supplied in a non-sterile state but packaged in such a way that it can be sterilised at a later stage following supply. This applies to any person or entity involved in the manufacture, importation, supply, or exportation of such medical devices within Australia. The Order sets out updated standards based on international benchmarks to ensure the quality, safety, and efficacy of medical devices. The geographic and jurisdictional reach of the Order is national, as it is made under the Therapeutic Goods Act 1989, which applies throughout Australia. The Order revokes and replaces the previous Medical Device Standards Order No. 3 – Medical Devices Required to be Sterile, which was made in 2003. Compliance with the standards set out in the Order is voluntary, and manufacturers may choose alternative means to demonstrate compliance with the essential principles. The Order may be extended or restricted through subordinate instruments, as determined by the Minister for Health and Ageing or their delegate.

Key Provisions

The Medical Device Standards Order (Standards for Medical Devices Required to be Sterile) 2008 (MDSO (Sterile Devices) 2008), made under section 41CB of the Therapeutic Goods Act 1989, introduces new standards for the manufacture of medical devices that are intended to be supplied in a sterile state. These standards are based on updated international standards for sterilisation and packaging, and they apply to both devices sterilised by the manufacturer and those supplied in a non-sterile state but packaged for later sterilisation (sections 1 and 2). The Order specifies that these standards must be applied in combination to achieve the final level of sterility (section 3). The Act requires that medical devices comply with certain essential principles regarding safety and performance. While compliance with the medical device standards set out in the Order is not mandatory, if a manufacturer applies these standards correctly, the device is presumed to comply with the relevant parts of the essential principles (section 41BH of the Act). The MDSO (Sterile Devices) 2008 sets out the requirements for sterile devices in Schedule 1 and for devices intended to be sterilised before supply in Schedule 2. Entities governed by the MDSO (Sterile Devices) 2008, such as manufacturers and importers of medical devices, are required to ensure their products meet the specified standards. This involves adhering to the latest international standards for sterilisation and packaging, which are referenced in the Order. Manufacturers must implement quality assurance techniques and validation processes as outlined in the relevant standards to ensure the sterility of their products. Failure to comply with the standards set out in the MDSO (Sterile Devices) 2008 may result in regulatory action under the Therapeutic Goods Act 1989. While compliance with the standards is voluntary, non-compliance may lead to scrutiny by the Therapeutic Goods Administration (TGA) to determine whether the essential principles are being met. If a device is found not to comply with the essential principles, it may be refused entry into Australia, recalled, or subject to enforcement actions such as fines or legal proceedings. The specific penalties for non-compliance are not detailed in the explanatory statement but are governed by the broader provisions of the Therapeutic Goods Act 1989, which may include substantial fines and imprisonment for serious or repeated breaches.